Cybird™ Gold : Dual Band LED Curing Light

DXM Co., Ltd. · #CGD-9000 · source

Cybird™ Gold : Dual Band LED Curing Light

Key Features
  • Maintains high intensity while generating low heat
  • Cures within 2 seconds (2mm depth materials)
  • 3 Curing Modes:
    • High Power Mode = 1,500 mW/cm² (5, 10, 15, 20 sec.)
    • Plasma Mode = 2,500mW/cm² (2, 3, 4 sec.)
    • Ortho Mode = 16 times of 3 sec. at 2,500 mW/cm²
  • Multi-wavelength allows for more compatibility with most resins (400nm ~ 490nm)
Components

The Cybird Gold includes light guides (45 and 90 degree configurations) with transilluminator tip and charger base.

Cleaning and Disinfection Instructions

Cleaning and Disinfection Instructions must be followed after each use.

Suggested Surface Disinfectant:

  • CaviCide® (or equivalent quaternary/alcohol blend disinfectant)

Do NOT Use:

  • Glutaraldehyde
  • Denatured alcohol
  • Lysol®
  • Phenol or phenolic cleaner
  • Ammonia complex
  • Iodine complex solutions

Disassembly and Inspection:

  1. Remove elastomer rubber light shield from glass fiber-optic light guide.
  2. Remove light guide from handpiece by gently pulling it straight out. Examine light guide ends for damage or composite material adhering to the tip. Examine light shield for tears or distortion. Replace if any damage is noted on either component.

Handpiece and Charger Base - Cleaning:

  1. Initial cleaning of handpiece must begin immediately after use to prevent drying of soil and contaminants in and on the device
  2. All exterior surfaces of the handpiece and charger base may be wiped with CaviWipes® or a cloth soaked in surface disinfectant to remove gross soil. Do not allow cleaning solution to invade the interior of these components since this may adversely affect the electronics.

Handpiece and Charger Base - Disinfecting:

  1. All exterior surfaces of the handpiece or charger base must be wiped and wetted with CaviWipes® or a cloth soaked in surface disinfectant.
  2. Allow the CaviCide® surface disinfectant to reside of the surface for a minimum of 3 minutes. Do not allow the disinfectant to dry on the surface.

Handpiece and Charger Base - Drying:

  1. Use a clean dry cloth to dry the exterior surfaces. Do not allow fluids to accumulate in the charger base socket as this may adversely affect the electronics.

Light Guide (Fiber Optic Glass) - Cleaning:

  1. Initial cleaning of the fiber optic light guide must begin immediately after use to prevent drying of soil and contaminants in and on the device.
  2. All exterior surfaces may be wiped with CaviWipes® or a cloth soaked in surface disinfectant to remove gross soil. Use a soft brush to remove contaminants in the junction between the glass rod and metal ferrule if necessary.
  3. Clean thoroughly using an ultrasonic cleaner. Minimum cycle time 10 minutes.

Light Guide (Fiber Optic Glass) - Rinsing:

  1. Using clean tap water on a clean cloth, wipe away residual disinfectant

Light Guide (Fiber Optic Glass) - Sterilizing: Package in an FDA cleared wrap prior to sterilization. Sterilization can be performed in gravity autoclave at 132° C / 270° F for 15 minutes with a 15-30 minute drying time.

Light Shield (Elastomer Rubber) - Cleaning:

  1. Initial cleaning of the light shield must begin immediately after use to prevent drying of soil and contaminants in and on the device.
  2. All exterior surfaces may be wiped with CaviWipes® or a cloth soaked in surface disinfectant to remove gross soil. Examine the light shield for damage and discard if any cuts, tears, or distortion is noted.
  3. Clean thoroughly using an ultrasonic cleaner. Minimum cycle time is 10 mins.
  4. The elastomer rubber light shield can be cleaned, sterilized and reused up to 5 times, discard after that.

Light Shield (Elastomer Rubber) - Rinsing:

  1. Using clean tap water on a clean cloth, wipe away residual disinfectant.

Light Shield (Elastomer Rubber) - Sterilizing: Package in a FDA cleared wrap prior to sterilization. Sterilization can be performed in a gravity autoclave at 132°C / 270°F for 15 mins. with a 15-30 min. drying time.

Reassembly & Storage:

  1. After all parts have been allowed to dry completely, gently insert the metal end of the glass fiber optic light guide into the circular opening of the handpiece. Make sure the light guide is completely inserted until it stops. This will position the light guide properly with the internal LED light source.
  2. Insert an elastomer rubber light shield over the end of the fiber optic light guide.
  3. Store the reassembled Cybird Curing light in a clean, dry location, preferably on the Cybird charger base so that the battery will be completely charged prior to the next use.

Product Information

ManufacturerDXM Co., Ltd.
Ortho Mode16 times of 3 sec. at 2,500 mW/cm^2
Wavelength400nm ~ 490nm
Plasma Mode2,500 mW/cm^2 (2, 3, 4 sec.)
Curing Modes3 Modes (High Power, Plasma, Ortho)
High Power Mode1,500 mW/cm^2 (5, 10, 15, 20 sec.)